Turbidity measuring point DULCOTEST DULCO turb C

Measuring range 0 – 1,000 NTU

Turbidity measurements with DULCOTEST DULCO turb C: Compact measuring instrument that uses light scatter to measure turbidity, with a large measuring range and different designs to comply with ISO and EPA standards. Available with or without automatic cleaning.

Product description

Reliable on-line measurement of turbidity with DULCOTEST DULCO turb C measuring points

The DULCOTEST measuring points for turbidity DULCO turb C with TUC 1, TUC 2, TUC 5, TUC 6 versions are compact, on-line turbidity measuring points consisting of a sensor, flow fitting and measuring instrument. The measuring instrument allows the measured value to be displayed, calibration When a metering pump has been installed, the capacity in litres per hour has to be determined under ... more in the glossary, the measured value to be forwarded using a 4– 20 mA signal and limit violations and equipment failure to be indicated. The measuring cuvette integrated in the measuring instrument allows the device to be operated in the process line bypass. The optical measuring equipment does not make contact with the measured medium.

The intended application is the treatment of potable water, in which DULCO turb C can be used in all treatment stages from raw water and filter monitoring to measurement of fine turbidity in dispensed potable water. Further applications include the monitoring of turbidity in slightly polluted process water, wastewater as well as water requiring treatment from the food and beverage industry up to turbidity values of 1,000 NTU. In contrast to the TUC 1/TUC 2 types, the measuring points TUC 5, TUC 6 are the successor models to types TUC 3 and TUC 4 and like these include an ultrasound-based self-cleaning function. When used in deposit-forming waters, this is particularly useful for extending the maintenance intervals. The new measuring instruments are virtually identical to the old devices. The only differences are that they are operated using 5 keys rather than 4 keys and feature a USB connector with the associated software functions:

  • Data capture and storage of 1 year’s worth of measurement and calibration When a metering pump has been installed, the capacity in litres per hour has to be determined under ... more in the glossary data
  • Variable recording intervals of between 1 and 60 minutes, which can be selected by the user
  • Downloading of data via a USB flash drive
  • Software updates via USB flash drive
  • 4-20 mA and RS-485 Modbus outputs at the same time

The measuring principle is similar to a scattered light measurement. The light beam radiated into the measuring cuvette filled with sample water is scattered on turbidity particles and the scattered light is measured at right angles (90°) to the radiated light (nephelometric measurement). The unit of measurement for turbidity can be given as a NTU (Nephelometric Turbidity Unit) or as an FNU (Formazin Nephelometric Unit). The measuring process in types TUC 1/TUC 5 (infrared light) corresponds to the global standard ISO 7027 and the European standard DIN EN 27027. The measuring process in types TUC 2/TUC 6 (white light) corresponds to the US standard USEPA 180.1.

Your benefits

  • Compact turbidity measuring station with integrated sensor, flow cuvette and measuring instrument saves space and is simple to install and operate.
  • High dynamic measuring range between 0.02 and 1,000 NTU permits broad-based use in all stages of potable water treatment. Also ideal for monitoring wastewater from clarification plants and for monitoring ruptures with filters.
  • Short response times thanks to small-volume measuring cuvette.
  • Long-term stable measurements, even in contaminated water, by the optional ultrasonic cleaning of the measuring cuvette.
  • Fast and simple calibration on-site by optionally available, pre-assembled and time-stable calibration standards.

Field of application

The following list of applications are general examples of generally suitable applications for this product. Please contact your local sales representative to talk about the applicable use of this product in your specific case / environment.

  • Potable water treatment, for all treatment steps: from raw water and filter monitoring to measuring fine turbidity in the potable water that is to be discharged
  • Monitoring of turbidity in slightly polluted industrial water, wastewater and water requiring treatment in the food and beverage industry up to a turbidity value of 1,000 NTU

    Technical Details

    • The measuring process in types TUC 1/TUC 5 (infrared light) corresponds to the global standard ISO 7027 and the European standard DIN EN 27027.
    • The measuring process in types TUC 2/TUC 6 (white light) corresponds to the US standard USEPA 180.1.

    Technical Data

    Measuring range

    a) TUC1, TUC4; factory calibrated: 0...1000 NTU, other measuring ranges: 0...10 NTU and 0...100 NTU
    b) TUC5, TUC6; factory-calibrated 0...100 NTU, other measuring ranges: 0...10 NTU and 0...1000 NTU

    Accuracy

    ± 2 % of the indicated value or ± 0.02 NTU below 40 NTU depending on which value is greater ± 5 % of the indicated value above 40 NTU

    Resolution

    0.0001 NTU below 10 NTU

    Response time

    Configurable

    Display

    Multiple row LCD display with background lighting

    Alarm relay

    Two programmable alarms, 120-240 VAC, 2 A Form C relay

    Output signal

    4…20 mA, 600 Ω, electric isolation: dual insulation, interference surge category II

    Communication interface

    Bi-directional RS-485, Modbus

    Max. pressure

    Integrated pressure regulator regulates
    TUC1, TUC2: 1380 kPa (200 psi)
    TUC 5, TUC6: 700 kPa (101,5 psi)
    based on the flow rate

    Flow

    6…60 l/h

    Temperature

    1…50 °C

    Materials in Contact With the Medium

    Polyamide (PA), silicone, polypropylene (PP), Viton®, stainless steel, borosilicate glass

    Voltage supply

    100 – 240 V AC, 47 – 63 Hz, 80 VA

    Hydraulic connections

    Black hose, inside 4.75 mm, outside 8 mm, installation in the bypass for the process main line

    Ambient conditions

    Not suitable for operation outdoors. Maximum operating altitude 2000 m above sea level. Maximum 95% relative air humidity (non-condensing).

    Standard

    Infrared light: ISO 7027, DIN EN 27027

    Dimensions H x W x D

    35 x 30 x 30 cm

    Shipping weight

    2.5 kg

    Standard

    Ultrasonic cleaning

    Order no.

    TUC 1

    Infrared light: ISO 7027, DIN EN 27027

    No

    1037696

    TUC 2

    White light: US EPA 180.1

    No

    1037695

    TUC 5

    Infrared light: ISO 7027, DIN EN 27027

    Yes

    1115440

    TUC 6

    White light: US EPA 180.1

    Yes

    1115441

    Downloads for Turbidity measuring point DULCOTEST DULCO turb C

    More in the Download Centre

    0 Results
    Filter by document type
    Application Example / Reference (0)
    Brochure / Flyer (0)
    Catalogue (0)
    Certificate / Declaration (0)
    Data Sheet (0)
    Drawing / Dimensional Drawing / 3D (0)
    Exploded View / Spare Parts (0)
    Material Safety Data Sheet (0)
    Operating Instructions (0)
    Others (0)
    Press Release (0)
    Publication / Article (0)
    Software (0)
    Wiring Diagram (0)

    Selected filters:

    Matching downloads

    Sort by Document type
    Load more matches ...
    No more matches available!

    Unfortunately, we did not find an appropriate match.

    All words written correctly?
    Search again with fewer or different search terms?

    Please try again later

    Service Unavailable

    The server is temporarily unable to service your request due to maintenance downtime or capacity problems. Please try again later.

    Overview

    Reliable on-line measurement of turbidity with DULCOTEST DULCO turb C measuring points

    The DULCOTEST measuring points for turbidity DULCO turb C with TUC 1, TUC 2, TUC 5, TUC 6 versions are compact, on-line turbidity measuring points consisting of a sensor, flow fitting and measuring instrument. The measuring instrument allows the measured value to be displayed, calibration When a metering pump has been installed, the capacity in litres per hour has to be determined under ... more in the glossary, the measured value to be forwarded using a 4– 20 mA signal and limit violations and equipment failure to be indicated. The measuring cuvette integrated in the measuring instrument allows the device to be operated in the process line bypass. The optical measuring equipment does not make contact with the measured medium.

    The intended application is the treatment of potable water, in which DULCO turb C can be used in all treatment stages from raw water and filter monitoring to measurement of fine turbidity in dispensed potable water. Further applications include the monitoring of turbidity in slightly polluted process water, wastewater as well as water requiring treatment from the food and beverage industry up to turbidity values of 1,000 NTU. In contrast to the TUC 1/TUC 2 types, the measuring points TUC 5, TUC 6 are the successor models to types TUC 3 and TUC 4 and like these include an ultrasound-based self-cleaning function. When used in deposit-forming waters, this is particularly useful for extending the maintenance intervals. The new measuring instruments are virtually identical to the old devices. The only differences are that they are operated using 5 keys rather than 4 keys and feature a USB connector with the associated software functions:

    • Data capture and storage of 1 year’s worth of measurement and calibration When a metering pump has been installed, the capacity in litres per hour has to be determined under ... more in the glossary data
    • Variable recording intervals of between 1 and 60 minutes, which can be selected by the user
    • Downloading of data via a USB flash drive
    • Software updates via USB flash drive
    • 4-20 mA and RS-485 Modbus outputs at the same time

    The measuring principle is similar to a scattered light measurement. The light beam radiated into the measuring cuvette filled with sample water is scattered on turbidity particles and the scattered light is measured at right angles (90°) to the radiated light (nephelometric measurement). The unit of measurement for turbidity can be given as a NTU (Nephelometric Turbidity Unit) or as an FNU (Formazin Nephelometric Unit). The measuring process in types TUC 1/TUC 5 (infrared light) corresponds to the global standard ISO 7027 and the European standard DIN EN 27027. The measuring process in types TUC 2/TUC 6 (white light) corresponds to the US standard USEPA 180.1.

    Your benefits

    • Compact turbidity measuring station with integrated sensor, flow cuvette and measuring instrument saves space and is simple to install and operate.
    • High dynamic measuring range between 0.02 and 1,000 NTU permits broad-based use in all stages of potable water treatment. Also ideal for monitoring wastewater from clarification plants and for monitoring ruptures with filters.
    • Short response times thanks to small-volume measuring cuvette.
    • Long-term stable measurements, even in contaminated water, by the optional ultrasonic cleaning of the measuring cuvette.
    • Fast and simple calibration on-site by optionally available, pre-assembled and time-stable calibration standards.

    Application

    Technical Data

    Downloads

    Contact us

    Wesley Sutton

    Managing Director, ProMinent Fluid Controls UK Ltd.

    »We are always available to help. Don't hesitate to contact us.«

    +44 1530 560 555  Send request   

    Matching Products

    ORP measurements with DULCOTEST sensors: Precise, reliable and application-adapted measured values in real-time. Can be used in water treatment or industrial process waters with strict requirements.

    more

    Chlorine dioxide measurement with DULCOTEST sensors offers precise, reliable, application-specific measured values in real-time for all oxidation and disinfection tasks.

    more

    The application-based DULCOTEST sensors for free chlorine deliver precise measured values and a high degree of monitoring and process reliability in any disinfection application.

    more

    Measuring ozone in oxidation and disinfection applications: DULCOTEST sensors offer precise, reliable, application-specific measured values in real-time.

    more

    Conductivity sensors for optimum process integration: DULCOTEST sensors meet a wide range of measuring requirements and allow the best solution to any given measuring task to be achieved.

    more

    ProMinent Newsletter - receive latest news, customized for you! subscribe now